ABB TB82PH Instrucciones de operaciones Pagina 28

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 60
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 27
I-E67-TB82FF-2A February 13, 2002
28
CONDUCTIVITY/CONCENTRATION TRANSDUCER BLOCK
IDX Name Description Bytes ST AR Default Valid Range Note
0 BLOCK_OBJ Block Object 62
1 ST_REV Static Revision 2 S R 0
2 TAG_DESC TAG Description 32 S R/W Spaces
3 STRATEGY Strategy 2 S R/W 0
4 ALERT_KEY Alert Key 1 S R/W 0 1 to 255
5 MODE_BLK Target 1 N R/W OOS Auto, OOS
Actual 1 D R
Permitted 1 S R/W Auto, OOS Auto, OOS
Normal 1 S R/W Auto Auto, OOS
6 BLOCK_ERR Block error 2 D R
0 – Off
1 – On
7 UPDATE_EVT Unacknowledged 1 D R/W 0
Update State 1 D R 0
Time Stamp 8 D R 0
Static Rev 2 D R 0
Relative Index 2 D R 0
8 BLOCK_ALM Unacknowledged 1 D R/W 0
Alarm State 1 D R 0
Time Stamp 8 D R 0
Sub-code 2 D R 0
Value 1 D R 0
9
TRANSDUCER_
DIRECTORY
Number of
Transducer
Directories &
Start Indexes
4NR
10
TRANSDUCER_
TYPE
2 N R 65535 65535 (Other)
11 XD_ERROR Transducer Error 1 D R 0
12
COLLECTION_
DIRECTORY
28 N R
13
PRIMARY_
VALUE_TYPE
2 S R/W 113
113– Conductivity
117–
Concentration
Note A
14
PRIMARY_
VALUE
Value 4 D R Note B
Status 1 D R
15
PRIMARY_
VALUE_RANGE
High Range 4 N R
1999000 µS/cm
Table 3 Note C
Low Range 4 N R
0 µS/cm
Table 3
Unit Index 2 N R
1586 (µS/cm)
Table 2
Decimal point 1 N R 2
16 CAL_UNIT Calibration Unit 2 S R/W
1586 (µS/cm)
Table 2
17 CAL_ONE_POINT
One-point Calibration
Value
4SR/W
0 µS/cm
Table 3
18 SENSOR_TYPE 2 S R/W 104 104 (ohms)
19
SENSOR_
RANGE
High Range 4 N R
1999000 µS/cm
Table 3
Low Range 4 N R
0 µS/cm
Table 3
Unit Index 2 N R
1586 (µS/cm)
Table 2
Decimal point 1 N R 2
Vista de pagina 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 59 60