I-E67-TB82FF-2A February 13, 2002
8
Property Characteristic/Value
Maximum Span
(full scale settings)
Conductivity
Concentration
Temperature
Sensor Group A: 1999 mS/cm
Sensor Group B: 1999 S/cm
Sensor Group C: 199.9 S/cm
1999 Digits
200
o
C, 360
o
F (0 to 200
o
C, 32 to 392
o
F)
Damping Continuously adjustable from 0.0 to 99.9 seconds
Supply Voltage 9 to 32 Vdc (9 to 24 Vdc for agency certified IS applications)
15 mA quiescent current
Turn-On Time 4 seconds typical, 6 seconds maximum
Maximum Sensor Cable Length 100 ft (30.5 m) Sensor Group A
50 ft (15.2 m) Sensor Group B
25 ft (7.6 m) Sensor Group C
Sensor Diagnostic Ground-Loop Detection, Dirty Sensor, Slope and Offset Check
Diagnostic Notification
Local HMI FAULT icon indication.
Environmental
Operating temperature
LCD Range
Storage temperature
-20
o
to 60
o
C (-4
o
to 140
o
F)
-20
o
to 60
o
C (-4
o
to 140
o
F)
-40
o
to 70
o
C (-40
o
to 158
o
F)
Mounting Effect None
Enclosure Classification NEMA 4X
IP65
Size
Height
Minimum panel depth
Maximum panel cutout
Recommended panel cutout
144 mm high x 144 mm wide x 171 mm long (5.66 in. high x 5.66 in. wide x
6.70 in. long)
145 mm (5.70 in.)
136.7 mm x 136.7 mm (5.38 in. x 5.38 in.)
135 mm x 135 mm (5.33 in. x 5.33 in.)
Weight 4.2 lb (1.9 kg) without mounting hardware
7.5 lb (3.4 kg) with Pipe Mounting Hardware
Comentarios a estos manuales