I-E67-TB82FF-2A February 13, 2002
2
OUT
OUT
Channel
SV2: Concentration or
Conductivity
SV1: Temperature
PV: Conductivity or
Concentration
Channel
Transducer
Block
Analog
Input
Function
Block 1
Analog
Input
Function
Block 2
Sensor
Input
Figure 33 - Possible Connections Between the TB and AI’s
The table below gives a summary of the relation between the selected measurement and available
variables/channels that can be used as an input for the Analog Input Function Blocks.
TYPE OF MEASURE (TB_PRIMARY_VALUE_TYPE)
Conductivity Concentration
Channel 1
(TB_PRIMARY_VALUE)
Conductivity Value Concentration Value
Channel 2
(TB_SECONDARY_VALUE_1)
Temperature Temperature
Channel 3
(TB_SECONDARY_VALUE_2)
Concentration Value Conductivity Value